SmartDrop™ X - Nano Spectrophotometer
The Accuris SmartDrop™ series of nano spectrophotometers provides quick and easy quantification of small samples. Utilizing a small sample volume eliminates the need for dilutions, providing more accurate results.
A full color, 7” touchscreen with icon-driven menus and preprogrammed protocols gets users up and running quickly. No external computer is required. Each unit features a quartz pedestal onto which a drop of sample is placed. The arm is lowered and the drop is stretched into a liquid column of specificed length for spectral analysis. DNA, RNA and protein concentrations and purity numbers are displayed in a matter of seconds.
The resulting data can be printed, saved in the SmartDrop’s™ onboard memory or exported via USB as an Excel file. All SmartDrop™ models include an integrated printer. For OD600 readings, each model includes a cuvette port and two optical glass cuvettes. Three SmartDrop models are available to meet a range of applications and budgets.

- Full-spectrum absorbance scans from 200 to 800nm
- Quickly calculates and displays 260/280 and 260/230 ratios
- Long-life Xenon lamp with precision monochromator
- Auto-ranging path lengths of 0.2mm and 1.0mm
- Detect up to 4,500 ng/µl of dsDNA without dilution
A full spectrum UV-Vis spectrophotometer, SmartDrop X requires 0.5 to 2.0µl of sample. In addition to nucleic acid and protein calculations, the software has preset programs for BCA, Bradford and Lowry assays. A full spectrum analysis can be performed in about 6 seconds, aiding in the
identification of unknowns. Up to 5 specific points on the spectrum can be selected for colorimetric readings output. The long life Xenon lamp source requires no warm up and is paired with a highly sensitive CCD detector. To provide the most accurate results, the SmartDrop X utilizes 2 path lengths.
SPECIFICATIONS
| Model |
SmartDrop L (NS1000) |
SmartDrop X (NS1010) |
SmartDrop XF (NS1020) |
| Display |
7-inch touch screen |
7-inch touch screen |
7-inch touch screen |
| Minimum sample size |
1.0µl – 2.0µl |
0.5µl – 2.0µl |
0.5µl – 2.0µl |
| Path length |
0.5mm |
0.2mm, 1.0mm |
0.05mm, 0.2mm, 1.0mm |
| Light source |
UV LED |
Xenon Flash Lamp |
Xenon Flash Lamp |
| Detector |
UV Silicon Photocell |
2048 Element Linear Silicon CCD Array |
2048 Element Linear Silicon CCD Array |
| Wavelength |
260, 280, & 600nm |
200 – 800nm (1nm steps) |
200 – 800nm (1nm steps) |
| Absorbance range |
0.2 – 50 A (10mm equivalent) |
0.04 – 90 A (10mm equivalent) |
0.04 – 300 A (10mm equivalent) |
| Absorbance precision |
0.005 A |
0.003 A |
0.003 A |
| Lower limit of detection |
dsDNA: 10ng/µl; BSA: 0.1 mg/ml |
dsDNA: 2ng/µl; BSA: 0.1mg/ml |
dsDNA: 2ng/µl ; BSA: 0.1mg/ml |
| Upper limit of detection |
dsDNA: 2,500ng/µl; BSA: 75.0 mg/ml |
dsDNA: 4,500ng/µl; BSA: 135mg/ml |
dsDNA: 15,000ng/µl; BSA: 360mg/ml |
| Detection time |
< 6 seconds |
< 5 seconds |
< 6 seconds |
| Communication ports |
USB for data export |
USB for data export |
USB for data export |
| Voltage input |
DC 24V, 2A |
DC 24V, 2A |
DC 24V, 2A |
| Power |
25W |
25W |
25W |
| Dimensions (WxDxH) |
8.2 x 11.0 x 7.3 in. / 20.8 x 28.0 x 18.6cm |
8.3 x 10.6 x 7.2 in. / 21.0 x 26.8 x 18.1cm |
8.2 x 12.6 x 7.3 in. / 20.8 x 32.0 x 18.6cm |
| Weight |
3.6 kg / 7.9 lbs |
2.8 kg / 6.2 lbs |
3.6 kg / 7.9 lbs |
| Cuvette OD600 detection |
|
|
|
| Absorbance range |
0.0 – 4.0 A |
0.0 – 4.0 A |
0.0 – 4.0 A |
| Absorbance stability |
[0,3) ≤ 0.5%, [3,4) ≤ 2% |
[0,3) ≤ 0.5%, [3,4) ≤ 2% |
[0,3) ≤ 0.5%, [3,4) ≤ 2% |
| Absorbance repeatibility |
[0,3) ≤ 0.5%, [3,4) ≤ 2% |
[0,3) ≤ 0.5%, [3,4) ≤ 2% |
[0,3) ≤ 0.5%, [3,4) ≤ 2% |
| Absorbance precision |
[0,2) ≤ 0.005 A, [2,3) ≤ 1%, [3,4) ≤ 2% |
[0,2) ≤ 0.005 A, [2,3) ≤ 1%, [3,4) ≤ 2% |
[0,2) ≤ 0.005 A, [2,3) ≤ 1%, [3,4) ≤ 2% |
| Fluorescence detection |
|
|
|
| Excitation/Emission |
N/A |
N/A |
460 ± 20nm / 525-570nm |
| Sensitivity |
N/A |
N/A |
dsDNA: 0.5pg/µl |
| Linearity |
N/A |
N/A |
R² ≥ 0.995 |
| Repeatability |
N/A |
N/A |
≤ 1.5% |
| Stability |
N/A |
N/A |
≤ 1.5% |